NaturalVue Enhanced Multifocal 1 Day
MultifocalMyopia ControlDaily
- BC
- 8.3
- Dia
- 14.5
- Dk
- 20
- H₂O
- 58%
- Power
- +4.00 to -12.25
Power Range
+4.00 to -12.25
CT @ −3.00D
0.08mm
Extended Wear
No
Material
etafilcon A
UV Blocking
HydrogelIonic

Availability● Available
Pack Sizes10, 30, 90
Available Powers
SPH: +4.00D to -12.25D (0.25D steps)
ADD: Universal (single add; no discrete add powers ordered)
Design Notes
Optics
Neurofocus Optics creates pinhole optics, which results in an extended depth of focus (EDOF) that delivers a wider range of clear vision along the visual axis.
Add
One universal design covers an effective add up to +3.00D, so no discrete add powers are ordered.
Fitting
Fit like a single-vision distance sphere lens; fitting success rate 88%.
Wetting agent
The TripleTear lubrication system uses three comfort agents, including hyaluronic acid, to provide hydration and lubrication, lock in moisture and maintain a clean surface.
Edge
An Ultra-Tapered Edge is designed for fit and comfort across all day wear.
Optical design
Range of clear vision is 76 cm.
Regulatory
Not FDA-approved for slowing myopia progression. US clearance is by 510(k), held by Visioneering Technologies (K150385, 2015-04-30 and K173086, 2017-11-07), for correction of refractive ametropia and presbyopia; FDA lists no PMA for this lens. An FDA myopia-control indication requires a PMA, as MiSight 1 day has (P180035).
Initial Fitting Guide
Refract to the sphero-cylindrical distance power that gives optimal vision; DO NOT push plus. Verify binocularly with the Duochrome (red-green) in a darkened room and reduce plus until the green-side letters are first seen clearer ("one click into green").
Initial lens
the one-click-into-green spherical equivalent DISTANCE Rx, vertexed to the corneal plane. VTI's QuickStart Calculator automates this step.
One universal add covers ADD needs to +3.00D; there is no add power to select. Get distance crisp and near follows.
| Complaint | Adjustment |
|---|---|
| Distance not crisp | Add -0.25D to the DOMINANT eye; a second -0.25D to the same eye if needed |
| Near not adequate | Add +0.25D to the NON-DOMINANT eye; a second +0.25D to the same eye if needed |
Over-refraction may not show a response; adjust from the patient's real-world vision reports rather than the phoropter. Full protocol: NaturalVue Professional Fitting Guide